Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Goerz SERVOGOR 790: (1 year) CAL VER /5520 DATE: 08-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 53 NUMBER OF LINES: 438 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # GOERZ SERVOGOR 790 Service Manual #331 0065 03 - 11/87 # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # Sub Goerz SERVOGOR 790 Linearity (1yr) CAL VER /5520 # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P J F W 1.002 ASK+ X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 30min - REQUIRED 1.004 DISP Temperature: 23 +/-2 degC 1.004 DISP Humidity: 55 +/-10% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP This procedure was designed to test 2 measuring units 1.006 DISP (1mV & .1mV). Since these units can be either the "X" 1.006 DISP or the "Y" inputs, the units will be identified as 1.006 DISP .1mV or 1mV (corresponding to max. sensitivity range). 1.007 DISP Install the proper chart paper and recording pen 1.007 DISP for proper operation of the UUT. 1.008 RSLT =CHANNEL Y TESTS 1.009 DISP Connect the 5520A to the UUT as follows: 1.009 DISP [32] 5520A NORMAL HI to Y INPUT HI 1.009 DISP [32] 5520A NORMAL LO to Y INPUT LO 1.009 DISP [32] 5520A AUX HI to X INPUT HI 1.009 DISP [32] 5520A AUX LO to X INPUT LO 1.010 OPBR Does the UUT have a 1mV Y input measuring unit? 1.011 JMPF 8.001 1.012 HEAD {Y INPUT: 1mV MEASURING UNIT} 1.013 HEAD {ZERO RANGE} 1.014 M5520 0V AX 1.015 5520 0mV S 2W 1.016 DISP Set S/CM switch on Time unit to X. 1.016 DISP Set Y Range to 1mV. 1.016 DISP Set Y CAL/VAR to CAL. 1.016 DISP Set X Range to 1mV. 1.016 DISP Set X CAL/VAR to CAL. 1.017 DISP Adjust the Y ZERO control for 0% vertical deflection 1.017 DISP on the chart paper. 1.017 DISP 1.017 DISP Adjust the X ZERO control for 50% horizontal deflection 1.017 DISP on the chart paper. 1.018 DISP Set OFF/CHART/PEN switch to PEN. 1.019 MESS Adjust the Y ZERO control through it's full range and 1.019 MESS verify that it is continously variable from -5% to 1.019 MESS 105% vertical deflection of the chart range. 1.019 MESS 1.020 EVAL Is {the zero adjustable from -5% to 105% of chart range}? 2.001 MESS 2.002 HEAD {DC VOLTAGE} 2.003 HEAD DC VOLTAGE: {1mV/cm Range} 2.004 5520 0V S 2W 2.005 DISP Adjust the Y ZERO control for 0% vertical deflection. 2.006 MESS Slew the 5520A for full scale deflection. 2.006 MESS 2.007 5520 1 18.0mV 0.5% 2W 3.001 HEAD DC VOLTAGE: {10mV/cm Range} 3.002 5520 0V S 2W 3.003 DISP Set Y Range to 10mV/cm. 3.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 3.005 5520 10 180.0mV 0.5% 2W 4.001 HEAD DC VOLTAGE: {.1V/cm Range} 4.002 5520 0V S 2W 4.003 DISP Set Y Range to .1V/cm. 4.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 4.005 5520 0.1 1.800V 0.5% 2W 5.001 HEAD DC VOLTAGE: {1V/cm Range} 5.002 5520 0V S 2W 5.003 DISP Set Y Range to 1V/cm. 5.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 5.005 5520 1 18.00V 0.5% 2W 6.001 HEAD DC VOLTAGE: {10V/cm Range} 6.002 5520 0V S 2W 6.003 DISP Set Y Range to 10V/cm. 6.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 6.005 5520 10 180.0V 0.5% 2W 7.001 MESS 7.002 M5520 * 7.003 MEM2 =Y 7.004 CALL Sub Goerz SERVOGOR 790 Linearity (1yr) CAL VER /5520 7.005 JMP 28.001 7.006 EVAL dummy 8.001 HEAD {Y INPUT: .1mV MEASURING UNIT} 8.002 HEAD {ZERO RANGE} 8.003 M5520 0V AX 8.004 5520 0mV S 2W 8.005 DISP Set S/CM switch on Time unit to X. 8.005 DISP Set Y mV/cm V/cm <0> switch to <0>. 8.005 DISP Set Y Range to .1mV. 8.005 DISP Set Y CAL/VAR to CAL. 8.005 DISP Set X Range to 1mV. 8.005 DISP Set X CAL/VAR to CAL. 8.006 DISP Adjust the Y ZERO control for 0% Vertical Deflection 8.006 DISP on the chart paper. 8.006 DISP 8.006 DISP Adjust the X ZERO control for 50% Horizontal Deflection 8.006 DISP on the chart paper. 8.007 DISP Set OFF/CHART/PEN switch to PEN. 8.008 MESS Adjust the Y ZERO control through it's full range and 8.008 MESS check that it is continously variable from -5% to 8.008 MESS 105% vertical deflection of the chart range. 8.008 MESS 8.009 EVAL Is {the zero adjustable from -5% to 105% of chart range}? 9.001 MESS 9.002 HEAD {DC VOLTAGE} 9.003 DISP Set Y mV/cm V/cm <0> switch to mV/cm. 9.004 HEAD DC VOLTAGE: {.1mV/cm Range} 9.005 5520 0V S 2W 9.006 DISP Adjust the Y ZERO control for 0% vertical deflection. 9.007 MESS Slew the 5520A for full scale deflection. 9.007 MESS 9.008 5520 0.1 1.800mV 0.5% 2W 10.001 HEAD DC VOLTAGE: {.2mV/cm Range} 10.002 5520 0V S 2W 10.003 DISP Set Y Range to .2mV/cm. 10.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 10.005 5520 0.2 3.600mV 0.5% 2W 11.001 HEAD DC VOLTAGE: {.5mV/cm Range} 11.002 5520 0V S 2W 11.003 DISP Set Y Range to .5mV/cm. 11.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 11.005 5520 0.5 9.00mV 0.5% 2W 12.001 HEAD DC VOLTAGE: {1mV/cm Range} 12.002 5520 0V S 2W 12.003 DISP Set Y Range to 1mV/cm. 12.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 12.005 5520 1 18.00mV 0.5% 2W 13.001 HEAD DC VOLTAGE: {2mV/cm Range} 13.002 5520 0V S 2W 13.003 DISP Set Y Range to 2mV/cm. 13.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 13.005 5520 2 36.00mV 0.5% 2W 14.001 HEAD DC VOLTAGE: {5mV/cm Range} 14.002 5520 0V S 2W 14.003 DISP Set Y Range to 5mV/cm. 14.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 14.005 5520 5 90.00mV 0.5% 2W 15.001 HEAD DC VOLTAGE: {10mV/cm Range} 15.002 5520 0V S 2W 15.003 DISP Set Y Range to 10mV/cm. 15.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 15.005 5520 10 180.0mV 0.5% 2W 16.001 HEAD DC VOLTAGE: {20mV/cm Range} 16.002 5520 0V S 2W 16.003 DISP Set Y Range to 20mV/cm. 16.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 16.005 5520 20 360.0mV 0.5% 2W 17.001 HEAD DC VOLTAGE: {50mV/cm Range} 17.002 5520 0V S 2W 17.003 DISP Set Y Range to 50mV/cm. 17.004 DISP Adjust the Y ZERO control for 0% vertical deflection. 17.005 5520 50 900.00mV 0.5% 2W 18.001 DISP Set mV/cm V/cm <0> switch to V/cm. 18.002 HEAD DC VOLTAGE: {.1V/cm Range} 18.003 5520 0V S 2W 18.004 DISP Set Y Range to .1V/cm. 18.005 DISP Adjust ZERO : 0% Vertical Deflection 18.006 5520 0.1 1.800V 0.5% 2W 19.001 HEAD DC VOLTAGE: {.2V/cm Range} 19.002 5520 0V S 2W 19.003 DISP Set Y Range to .2V/cm. 19.004 DISP Adjust ZERO : 0% Vertical Deflection 19.005 5520 0.2 3.600V 0.5% 2W 20.001 HEAD DC VOLTAGE: {.5V/cm Range} 20.002 5520 0V S 2W 20.003 DISP Set Y Range to .5V/cm. 20.004 DISP Adjust ZERO : 0% Vertical Deflection 20.005 5520 0.5 9.00V 0.5% 2W 21.001 HEAD DC VOLTAGE: {1V/cm Range} 21.002 5520 0V S 2W 21.003 DISP Set Y Range to 1V/cm. 21.004 DISP Adjust ZERO : 0% Vertical Deflection 21.005 5520 1 18.00V 0.5% 2W 22.001 HEAD DC VOLTAGE: {2V/cm Range} 22.002 5520 0V S 2W 22.003 DISP Set Y Range to 2V/cm. 22.004 DISP Adjust ZERO : 0% Vertical Deflection 22.005 5520 2 36.00V 0.5% 2W 23.001 HEAD DC VOLTAGE: {5V/cm Range} 23.002 5520 0V S 2W 23.003 DISP Set Y Range to 5V/cm. 23.004 DISP Adjust ZERO : 0% Vertical Deflection 23.005 5520 5 90.00V 0.5% 2W 24.001 HEAD DC VOLTAGE: {10V/cm Range} 24.002 5520 0V S 2W 24.003 DISP Set Y Range to 10V/cm. 24.004 DISP Adjust ZERO : 0% Vertical Deflection 24.005 5520 10 180.0V 0.5% 2W 25.001 HEAD DC VOLTAGE: {20V/cm Range} 25.002 5520 0V S 2W 25.003 DISP Set Y Range to 20V/cm. 25.004 DISP Adjust ZERO : 0% Vertical Deflection 25.005 MESS Slew the 5520A for 50% scale deflection. 25.005 MESS 25.006 5520 20 180.0V 0.5% 2W 26.001 HEAD DC VOLTAGE: {50V/cm Range} 26.002 5520 0V S 2W 26.003 DISP Set Y Range to 50V/cm. 26.004 DISP Adjust ZERO : 0% Vertical Deflection 26.005 MESS Slew the 5520A for 25% scale deflection. 26.005 MESS 26.006 5520 50 225.00V 0.5% 2W 27.001 MESS 27.002 M5520 * 27.003 MEM2 =Y 27.004 CALL Sub Goerz SERVOGOR 790 Linearity (1yr) CAL VER /5520 27.005 JMP 28.001 27.006 EVAL dummy 28.001 DISP This procedure was designed to test 2 measuring units 28.002 DISP (1mV & .1mV). Since these units can be either the "X" 28.002 DISP or the "Y" inputs, the units will be identified as 28.002 DISP .1mV or 1mV (corresponding to max. sensitivity range). 28.003 RSLT =CHANNEL X TESTS 28.004 DISP Connect the 5520A to the UUT as follows: 28.004 DISP [32] 5520A NORMAL HI to X INPUT HI 28.004 DISP [32] 5520A NORMAL LO to X INPUT LO 28.004 DISP [32] 5520A AUX HI to Y INPUT HI 28.004 DISP [32] 5520A AUX LO to Y INPUT LO 28.005 OPBR Does the UUT have a 1mV X input measuring unit? 28.006 JMPF 34.006 28.007 HEAD {X INPUT: 1mV MEASURING UNIT} 28.008 HEAD {ZERO RANGE} 28.009 M5520 0V AX 28.010 5520 0mV S 2W 28.011 DISP Set S/CM switch on Time unit to X. 28.011 DISP Set X Range to 1mV. 28.011 DISP Set X CAL/VAR to CAL. 28.011 DISP Set Y Range to 1mV. 28.011 DISP Set Y CAL/VAR to CAL. 28.012 DISP Adjust the X ZERO control for 0% horizontal deflection 28.012 DISP on the chart paper. 28.012 DISP 28.012 DISP Adjust the Y ZERO control for 50% vertical deflection 28.012 DISP on the chart paper. 28.013 DISP Set OFF/CHART/PEN switch to PEN. 28.014 MESS Adjust the X ZERO control through it's full range and 28.014 MESS check that it is continously variable from -5% to 28.014 MESS 105% horizontal deflection of the chart range. 28.014 MESS 28.015 EVAL Is {the zero adjustable from -5% to 105% of chart range}? 29.001 MESS 29.002 HEAD {DC VOLTAGE} 29.003 HEAD DC VOLTAGE: {1mV/cm Range} 29.004 5520 0V S 2W 29.005 DISP Adjust the X ZERO control for 0% horizontal deflection. 29.006 MESS Slew the 5520A for full scale deflection. 29.006 MESS 29.007 5520 1 28.0mV 0.5% 2W 30.001 HEAD DC VOLTAGE: {10mV/cm Range} 30.002 5520 0V S 2W 30.003 DISP Set X Range to 10mV/cm. 30.004 DISP Adjust the X ZERO control for 0% horizontal deflection. 30.005 5520 10 280.0mV 0.5% 2W 31.001 HEAD DC VOLTAGE: {.1V/cm Range} 31.002 5520 0V S 2W 31.003 DISP Set X Range to .1V/cm. 31.004 DISP Adjust the X ZERO control for 0% horizontal deflection. 31.005 5520 0.1 2.800V 0.5% 2W 32.001 HEAD DC VOLTAGE: {1V/cm Range} 32.002 5520 0V S 2W 32.003 DISP Set X Range to 1V/cm. 32.004 DISP Adjust the X ZERO control for 0% horizontal deflection. 32.005 5520 1 28.00V 0.5% 2W 33.001 HEAD DC VOLTAGE: {10V/cm Range} 33.002 5520 0V S 2W 33.003 DISP Set X Range to 10V/cm. 33.004 DISP Adjust the X ZERO control for 0% horizontal deflection. 33.005 MESS Slew the 5520A for 50% scale deflection. 33.005 MESS 33.006 5520 10 140.0V 0.5% 2W 34.001 MESS 34.002 M5520 * 34.003 MEM2 =X 34.004 CALL Sub Goerz SERVOGOR 790 Linearity (1yr) CAL VER /5520 34.005 JMP 53.005 34.006 HEAD {X INPUT: .1mV MEASURING UNIT} 34.007 HEAD {ZERO RANGE} 34.008 M5520 0V AX 34.009 5520 0mV S 2W 34.010 DISP Set S/CM switch on Time unit to X. 34.010 DISP Set X Range to .1mV. 34.010 DISP Set X CAL/VAR to CAL. 34.010 DISP Set Y Range to 1mV. 34.010 DISP Set Y CAL/VAR to CAL. 34.010 DISP Set X input as follows: 34.010 DISP Set X mV/cm V/cm <0> switch to <0>. 34.011 DISP Adjust the X ZERO control for 0% horizontal deflection 34.011 DISP on the chart paper. 34.011 DISP 34.011 DISP Adjust the Y ZERO control for 50% vertical deflection 34.011 DISP on the chart paper. 34.012 DISP Set OFF/CHART/PEN switch to PEN. 34.013 MESS Adjust the X ZERO control through it's full range and 34.013 MESS check that it is continously variable from -5% to 34.013 MESS 105% horizontal deflection of the chart range. 34.013 MESS 34.014 EVAL Is {the zero adjustable from -5% to 105% of chart range}? 35.001 MESS 35.002 HEAD {DC VOLTAGE} 35.003 DISP Set X mV/cm V/cm <0> switch to mV/cm. 35.004 HEAD DC VOLTAGE: {.1mV/cm Range} 35.005 5520 0V S 2W 35.006 DISP Adjust the X ZERO control to 0% horizontal deflection. 35.007 MESS Slew the 5520A for full scale deflection. 35.007 MESS 35.008 5520 0.1 2.800mV 0.5% 2W 36.001 HEAD DC VOLTAGE: {.2mV/cm Range} 36.002 5520 0V S 2W 36.003 DISP Set X Range to .2mV/cm 36.004 DISP Adjust the X ZERO to 0% horizontal deflection. 36.005 5520 0.2 5.600mV 0.5% 2W 37.001 HEAD DC VOLTAGE: {.5mV/cm Range} 37.002 5520 0V S 2W 37.003 DISP Set X Range to .5mV/cm 37.004 DISP Adjust the X ZERO to 0% horizontal deflection. 37.005 5520 0.5 14.00mV 0.5% 2W 38.001 HEAD DC VOLTAGE: {1mV/cm Range} 38.002 5520 0V S 2W 38.003 DISP Set X Range to 1mV/cm 38.004 DISP Adjust the X ZERO to 0% horizontal deflection. 38.005 5520 1 28.00mV 0.5% 2W 39.001 HEAD DC VOLTAGE: {2mV/cm Range} 39.002 5520 0V S 2W 39.003 DISP Set X Range to 2mV/cm 39.004 DISP Adjust the X ZERO to 0% horizontal deflection. 39.005 5520 2 56.00mV 0.5% 2W 40.001 HEAD DC VOLTAGE: {5mV/cm Range} 40.002 5520 0V S 2W 40.003 DISP Set X Range to 5mV/cm 40.004 DISP Adjust the X ZERO to 0% horizontal deflection. 40.005 5520 5 140.00mV 0.5% 2W 41.001 HEAD DC VOLTAGE: {10mV/cm Range} 41.002 5520 0V S 2W 41.003 DISP Set X Range to 10mV/cm. 41.004 DISP Adjust the X ZERO to 0% horizontal deflection. 41.005 5520 10 280.0mV 0.5% 2W 42.001 HEAD DC VOLTAGE: {20mV/cm Range} 42.002 5520 0V S 2W 42.003 DISP Set X Range to 20mV/cm. 42.004 DISP Adjust the X ZERO to 0% horizontal deflection. 42.005 5520 20 560.0mV 0.5% 2W 43.001 HEAD DC VOLTAGE: {50mV/cm Range} 43.002 5520 0V S 2W 43.003 DISP Set X Range to 50mV/cm. 43.004 DISP Adjust the X ZERO to 0% horizontal deflection. 43.005 5520 50 1400.00mV 0.5% 2W 44.001 DISP Set X mV/cm V/cm <0> switch to V/cm. 44.002 HEAD DC VOLTAGE: {.1V/cm Range} 44.003 5520 0V S 2W 44.004 DISP Set X Range to .1V/cm. 44.005 DISP Adjust the X ZERO to 0% horizontal deflection. 44.006 5520 0.1 2.800V 0.5% 2W 45.001 HEAD DC VOLTAGE: {.2V/cm Range} 45.002 5520 0V S 2W 45.003 DISP Set X Range to .2V/cm. 45.004 DISP Adjust the X ZERO to 0% horizontal deflection. 45.005 5520 0.2 5.600V 0.5% 2W 46.001 HEAD DC VOLTAGE: {.5V/cm Range} 46.002 5520 0V S 2W 46.003 DISP Set X Range to .5V/cm. 46.004 DISP Adjust the X ZERO to 0% horizontal deflection. 46.005 5520 0.5 14.00V 0.5% 2W 47.001 HEAD DC VOLTAGE: {1V/cm Range} 47.002 5520 0V S 2W 47.003 DISP Set X Range to 1V/cm. 47.004 DISP Adjust the X ZERO to 0% horizontal deflection. 47.005 5520 1 28.00V 0.5% 2W 48.001 HEAD DC VOLTAGE: {2V/cm Range} 48.002 5520 0V S 2W 48.003 DISP Set X Range to 2V/cm. 48.004 DISP Adjust the X ZERO to 0% horizontal deflection. 48.005 5520 2 56.00V 0.5% 2W 49.001 HEAD DC VOLTAGE: {5V/cm Range} 49.002 5520 0V S 2W 49.003 DISP Set X Range to 5V/cm. 49.004 DISP Adjust the X ZERO to 0% horizontal deflection. 49.005 5520 5 140.00V 0.5% 2W 50.001 HEAD DC VOLTAGE: {10V/cm Range} 50.002 5520 0V S 2W 50.003 DISP Set X Range to 10V/cm. 50.004 DISP Adjust the X ZERO to 0% horizontal deflection. 50.005 MESS Slew the 5520A for 50% scale deflection. 50.005 MESS 50.006 5520 10 140.0V 0.5% 2W 51.001 HEAD DC VOLTAGE: {20V/cm Range} 51.002 5520 0V S 2W 51.003 DISP Set X Range to 20V/cm. 51.004 DISP Adjust the X ZERO to 0% horizontal deflection. 51.005 MESS Slew the 5520A for 25% scale deflection. 51.005 MESS 51.006 5520 20 140.0V 0.5% 2W 52.001 HEAD DC VOLTAGE: {50V/cm Range} 52.002 5520 0V S 2W 52.003 DISP Set X Range to 50V/cm. 52.004 DISP Adjust the X ZERO to 0% horizontal deflection. 52.005 MESS Slew the 5520A for 10% scale deflection. 52.005 MESS 52.006 5520 50 140.00V 0.5% 2W 53.001 MESS 53.002 M5520 * 53.003 MEM2 =X 53.004 CALL Sub Goerz SERVOGOR 790 Linearity (1yr) CAL VER /5520 53.005 END